Appendix RIGOL
DL3000 Performance Verification Manual
15
Resistance Readback
Model
Range
Resolution
DL3021
0.08 Ω~15 kΩ
(0.0667 S~0.5 S)
2 mA/Vsense
DL3021A
DL3031
0.08 Ω~15 kΩ
(0.0667 S~0.5 S)
2 mA/Vsense
DL3031A
Power Readback
Model
Range
Resolution
DL3021
0 ~ 200 W
100 mW
DL3021A
DL3031
0 ~ 350 W
100 mW
DL3031A
Protection Function
Overcurrent protection (OCP), overvoltage protection (OVP), overpower protection
(OPP), overtemperature protection (OTP), as well as local/remote reverse voltage
(LRV/RRV) protection.
Stability
[7]
Model
Current
Voltage
DL3021
± (0.01%± 10 mA)
± (0.01%± 10 mV)
DL3021A
DL3031
± (0.01%± 10 mA)
± (0.01%± 10 mV)
DL3031A
Input Resistance
350 kΩ
Mechanical
Dimensions
239 mm(W) x 157 mm(H) x 442 mm(D)
Weight
Net weight: 7.58 kg
Power
AC Input (50 Hz~60 Hz)
115 Vac± 10%, 230 Vac± 10% (max: 250 Vac)
Maximum Input Power
<30 VA
Interface
USB Device Interface
1
USB HOST
1
LAN Interface
1 (optional)

Other manuals for Rigol DL3000

Related product manuals