Chapter 2 Constant Voltage Tests RIGOL
DP800 Performance Verification Manual
2-21
accordance with the value selected by the voltage selector at the rear panel of
DP800.
3. Measure the CV programming and readback accuracy (take CH1 as an example
in this place) when the output voltage is 0V.
1) Turn on DP800. Send the following commands via the remote interface to
select the channel to be tested (namely the current channel), set the
voltage and current of this channel (please refer to Table 2-22) as well as
enable the output of this channel.
:APPL CH1,0,5.3
/*Select CH1 as the current channel and set the
voltage and current of this channel to 0V and
5.3A respectively*/
:OUTP CH1,ON
/*Enable the output of CH1*/
Notes:
The above commands are only applicable to the multi-channel models
of DP800. You can also send the :INST:NSEL 1 (or :INST CH1)
command to select CH1 as the current channel, the :SOUR1:VOLT 0
command to set the voltage of CH1 to 0V and the :SOUR1:CURR 5.3
command to set the current of CH1 to 5.3A.
For the single channel model with multiple ranges of DP800 series
(DP811A), you can send the :OUTP:RANG P20V (or :OUTP:RANG LOW,
or :VOLT:RANG P20V, or :VOLT:RANG LOW) command to select Range1
as the current range and then send the :VOLT 0 command to set the
voltage of the channel to 0V and the :CURR 10.5 command to set the
current of the channel to 10.5A.
2) Turn on the digital multimeter; select DC voltage measurement function
and set the range to Auto. Make sure that the power supply is in CV mode.
Read and record the reading (U
1
) of the multimeter.
3) Send the :MEAS:VOLT? CH1 command via the remote interface; read and
record the returned voltage value U
2
of the query command.
4) Calculate the CV programming accuracy (|U
1
-0|) and the CV readback
accuracy (|U
2
-U
1
|). Compare them with the specified specifications in Table
2-22.
4. Measure the CV programming and readback accuracy (take CH1 as an example
in this place) when the output voltage is the rated output value.
1) Send the following command via the remote interface to select the channel
to be tested (namely the current channel) and set the voltage and current

Other manuals for Rigol DP831A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP831A and is the answer not in the manual?

Rigol DP831A Specifications

General IconGeneral
BrandRigol
ModelDP831A
CategoryPower Supply
LanguageEnglish

Related product manuals