EasyManua.ls Logo

Rigol DP831A - Page 72

Rigol DP831A
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Appendix
DP800 Performance Verification Manual
4
CC Programming and Readback Accuracy
Channel CH1 (8V/5A) CH2 (30V/2A) CH3 (-30V/2A)
Specification
Programming <20mA <9mA <9mA
Readback <20mA <7mA <7mA
The output current is 0
DP831A
Setting
Voltage 8.4V 32V -32V
Current 0A 0A 0A
Measurement
Result
U
1
I
1
=U
1
/R
M
[2]
I
2
Programming:
|I
1
-0|
Readback:
|I
2
-I
1
|
The output current is I
R
[3]
DP831A
Setting
Voltage 8.4V 32V -32V
Current 5A 2A 2A
Measurement
Result
U
3
I
3
=U
3
/R
M
[2]
I
4
Programming:
|I
3
-I
R
[3]
|
Readback:
|I
4
-I
3
]
|
Note
[1]
: The rated voltage of the specified channel.
Note
[2]
: R
M
=0.022Ω.
Note
[3]
: The rated current of the specified channel.

Other manuals for Rigol DP831A

Related product manuals