Chapter 2 Performance Verification Test RIGOL
DSA1000A/DSA1000 Performance Verification Guide
2-7
Absolute Amplitude Accuracy Test
Specification
Absolute Amplitude Accuracy
fc = 50 MHz, peak detector, preamplifier off, 10 dB attenuation, input signal = -10 dBm, 20 to 30
Uncertainty ±0.4 dB
Test Devices
1. Signal Generator × 1
2. Power Meter × 1
3. Power Sensor × 1
4. Power Meter Connecting Cable × 1
5. Dual-N Cable × 1
6. Dual-BNC Cable × 1
Test Connection Diagram
Figure 2-3 Absolute Amplitude Accuracy Test Connection Diagram