DSA1000A/DSA1000 Performance Verification Guide
5
10 MHz Reference Output Accuracy Test:
Frequency Counter
Measurement Result
Output Accuracy
[1]
Limit Pass/Fail
< 10 MHz*(3 ppm × time from the last
calibration till now + 3 ppm)
Note: [1] Output Accuracy = |Frequency Counter Measurement Result 10 MHz|