EasyManua.ls Logo

Rigol RSA Series - Page 56

Rigol RSA Series
89 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Chapter 2 Functions of the Front Panel
2-36 VNA User Guide
Table 2-38 Stop Distance
Parameter
Remarks
Default
6.72 m
Range
[1]
0.231 m to 150 Gm
Unit
Meter: m, dm, cm
Feet: ft
Knob Step
Stop distance/100
Left/Right Arrow Key Step
Up/Down Arrow Key Step
Stop distance/10
Note
[1]
: The values of Stop Distance, Span, Sweep Points, and Velocity Factor are
correlative. The formula is shown as follows:
 =
󰇛
 1
󰇜
 1.5 10
8

Remarks:
Once the value of Stop Distance is modified, the value of Sweep Points will be
automatically modified.
Port Velocity Factor
Sets the Velocity Factor of cable under test relative to the velocity of light in the
vacuum. The velocity factor for the common dielectric materials is as follows:
VF (velocity factor) for solid PE (solid polyethylene) is about 0.66.
VF (velocity factor) for solid PTFE (Polytetrafluoroethylene) is about 0.70.
Table 2-39 Velocity Factor
Parameter
Remarks
Default
0.66
Range
0.1 to 1
Unit
N/A
Knob Step
0.1
Left/Right Arrow Key Step
Up/Down Arrow Key Step
at 1-2-5 step
Port Specific Loss
Sets the cable loss for the port. It is used to compensate the attenuation of the
stimulus signal in different positions of a cable.

Related product manuals