7-5. Parts replacement
7-5-1. Periodic replacement parts
The consumable parts of the product are listed below. Consumable parts should be replaced using the
recommended replacement intervals as a guide.
<Recommended replacement parts list>
Name
Recommended
check interval
Recommended
replacement
interval
Quantity
(items
per unit)
Remarks
Combustible gas
sensor (NCR-6309)
6 months 3 years 1 *
O
2
sensor (ESR-X13P)
6 months 3 years 1 *
CO/H
2
S sensor
(ESR-A1DP)
6 months 3 years 1 *
CO sensor (ESR-A13P)
6 months 3 years 1 *
CO sensor (ESR-A1CP)
6 months 3 years 1 *
H
2
S sensor (ESR-A13i)
6 months 3 years 1 *
Dust filter
Before and
after use
6 months or
when contaminated
1 Part No.: 4777 9345 60
Interference gas
removal filter
3 months 6 months 1
For combustible gas sensor
(NCR-6309)
Part No.: 4777 9212 30
Interference gas
removal filter
3 months 6 months 1
For CO/H
2
S sensor
(ESR-A1DP)
Part No.: 4777 9314 10
(ESR-A13i)
Part No.: 4777 9317 30
Interference gas
removal filter
3 months 6 months 1
For CO sensor
(ESR-A1CP, ESR-A13P)
Part No.: 4777 9213 10
Interference gas
removal filter
3 months 6 months 1
For H
2
S sensor (ESR-A13i)
Part No.: 4777 9214 80
Rubber seals
- 3 to 6 years
1 set *
Battery
-
Approx. 500
charging/discharging
cycles
1 *
* A functional check by a qualified service engineer is required after replacement. To ensure safety and the
stable operation of the product, request checking by a qualified service engineer. Contact Riken Keiki to
request checking.
NOTE
The above replacement intervals are guidelines only. Replacement intervals may vary depending on
actual operating conditions. These intervals do not constitute warranty periods. Replacement
intervals may vary depending on the results of regular maintenance.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Riken Keiki GX-3R and is the answer not in the manual?

Riken Keiki GX-3R Specifications

General IconGeneral
Detection GasesLEL, O2, H2S, CO
Detection PrincipleCatalytic combustion (LEL), Electrochemical (O2, H2S, CO)
LEL Range0-100% LEL
BatteryRechargeable lithium-ion battery
Humidity Range0 to 95% RH (non-condensing)
Device TypeMulti-gas monitor
Target GasesOxygen, Hydrogen Sulfide, Carbon Monoxide
O2 Range0-30% vol
H2S Range0 to 100 ppm
Alarm Set PointsAdjustable
Alarm MethodAudible, visual, and vibration alarms
DisplayLCD with backlight
Operating Temperature-20 to +50°C
Protection degreeIP66
CO Range0-500 ppm