PT1E-1099
Indicator/Alarm Unit
RM-6000 Series
Operating Manual

Questions and Answers

  • P
    Patricia WallaceAug 13, 2025
    What to do if Riken Keiki Security System has abnormal operations?
    • B
      brian04Aug 13, 2025
      To address abnormal operations in your Riken Keiki Security System, first turn off and then restart the indicator/alarm unit. If this issue occurs frequently, take steps to eliminate any potential sources of noise.
  • R
    Rachel MolinaAug 17, 2025
    Why is my Riken Keiki OX-6002 responding slowly?
    • M
      Miranda RandallAug 17, 2025
      To resolve a slow response issue with your Riken Keiki Security System, start by replacing the dust filter if it's clogged. Inspect the suction or exhaust tubes for any bends or clogs and fix them accordingly. Also, check for any condensation inside the suction tube and rectify any issues. Finally, if the sensor sensitivity has deteriorated, replace the sensor unit with a new one.
  • D
    Dr. Melissa HortonAug 21, 2025
    Why span adjustment is impossible in Riken Keiki OX-6002 Security System?
    • D
      Dennis WalshAug 21, 2025
      If you are unable to perform span adjustment on your Riken Keiki Security System, verify that you are using the correct calibration gas. If the issue persists, the sensor sensitivity may have deteriorated, requiring a sensor replacement.
  • D
    David DavisAug 26, 2025
    What to do if the reading rises (drops) on Riken Keiki OX-6002?
    • D
      dustinhuntAug 26, 2025
      If the reading on your Riken Keiki Security System rises or drops and remains that way, perform a zero adjustment (fresh air adjustment) to correct for sensor output drift. If interference gas is present, contact Riken Keiki for information on actions, such as removal filter. Address any slow leaks by taking actions similar to those for a gas alarm. Also, perform a zero adjustment to compensate for environmental changes.
  • K
    Kathy RobinsonAug 29, 2025
    Why Riken Keiki OX-6002 triggers gas alarm with no leak?
    • B
      Brian SanfordAug 29, 2025
      If your Riken Keiki Security System triggers a gas alarm without a gas leak or other abnormalities, it could be due to interference gas. In this case, contact Riken Keiki for information on actions, such as removal filter. Another cause might be disturbance by noise; try turning off and restarting the indicator/alarm unit, and if the issue persists, take measures to eliminate the noise. A sudden change in the environment may also be the reason and you must take any preventive actions to eliminate them.
  • J
    Jennifer ReyesSep 1, 2025
    How to fix cable open-circuit in Riken Keiki OX-6002 Security System?
    • H
      Hannah JonesSep 1, 2025
      To fix a cable open-circuit between the indicator/alarm unit and the detector head of your Riken Keiki Security System, restore the cable connection.
  • C
    Christopher HarrellSep 5, 2025
    What causes sensor abnormalities E-1A SENSOR in Riken Keiki OX-6002 Security System?
    • M
      Mary BuchananSep 5, 2025
      To address sensor abnormalities E-1A SENSOR on your Riken Keiki Security System (GP-6001/NC-6001/GH-6001/OX-6001 only), check the harness connections between the amplifier PCB and the main PCB. If there's a sensor open-circuit or short-circuit in the detector head, properly connect the cable between the indicator/alarm unit and the detector head (sensor), or replace the sensor. For abnormalities in current supply to the sensor, readjust the heater current. If there are abnormalities in voltage supply, readjust the load voltage. Finally, for zero follower abnormalities, perform the zero adjustment.
  • L
    lgomezSep 9, 2025
    What to do if span adjustment is impossible in Riken Keiki OX-6002 Security System?
    • J
      Joseph WilliamsSep 9, 2025
      If you are unable to perform span adjustment on your Riken Keiki Security System, verify that you are using the proper calibration gas. If the issue persists, the sensor sensitivity may have deteriorated, requiring a sensor replacement.

Summary

Important Notices on Safety

Danger Cases

Lists critical safety warnings, e.g., unit is not explosion-proof.

Warning Cases

Outlines general warnings regarding power supply, grounding, and operation environment.

Operational Precautions

Provides specific operational cautions, e.g., avoiding radio interference and proper restart.

Product Components

How to Use

Wiring Procedures

Explains recommended cables and connection methods for various models.

How to Operate

Operations and Functions

Gas Alarm Activation

Explains how gas alarms are triggered, indicated, and their self-latching behavior.

Fault Alarm Activation

Details how the unit detects faults and displays error messages.

External Output Operation

Describes the 4-20mA and RS-485 signal outputs and their behavior.

Maintenance

Maintenance Intervals and Items

Lists recommended maintenance tasks and their frequencies.

Regular Maintenance Mode

Procedures for performing detailed maintenance and adjustments.

Gas Calibration Method

Steps for zero and span adjustments using calibration gases.

Troubleshooting

Unit Abnormalities

Lists common unit malfunctions, their causes, and recommended actions.

Product Specifications

Riken Keiki OX-6002 Specifications

General IconGeneral
Gas TypeOxygen (O2)
Operating Humidity95%RH (non-condensing)
Alarm MethodVisual and audible alarms
DisplayLCD display
CertificationsATEX, IECEx
Resolution0.1 vol%
Alarm Set PointsAdjustable