Communications Manual Rev 3.10  Software Versions 3.xx 
Page 14 
003R-618-310
 
4.  Protocol Overview 
The instrument contains a number of registers.  The indicator is configured by reading and 
writing  the  information  stored  in  these  registers.    Information  such  as  gross  weight  is 
obtained by reading these registers. The instrument has registers of different types to hold 
weights, menu options, etc. 
4.1.  COMM Message Structure 
The messages use ASCII characters.  Almost all numeric values are transmitted in 
hexadecimal form using uppercase HEX characters  (ie. 0 to 9,  A to F). 
Commands from a Master (typically a PC) to a Slave (typically an Indicator) are of 
the form: 
Address 
Field 
Command 
Field 
Register Id    [Optional 
Parameter] 
 
xx  mm  rrrr  “:”  [ppppp] 
  
Table 2 COMM Command Format 
Responses from Slave to Master are of the form: 
Address 
Field 
Command 
Field 
Register Id    Return value   
yy  mm  rrrr  “:”  vvvvv 
  
Table 3 COMM Response Format 
Where: 
xx is the Address Field sent to the indicator (Response bit = 0). 
yy is the Address Field in the response from the indicator (Response bit = 1). 
mm is the Command Field. 
rrrr is the Register Id Field. 
“:” is the separator between the command header and the parameter or return 
value. 
ppppp is the optional parameter value sent to the indicator. 
vvvvv is the return value from the indicator. 
  is the two characters <CR><LF> (Carriage-Return, Line-Feed).