Rinstrum -  R300 Series Digital Indicator Quick Start Manual Rev 1.5 
Page 18 
7. Instrument Setup  
7.1. Calibration Counter 
The R300 Series built-in calibration counter(s) monitor the 
number of times trade critical settings are altered.  Refer to Trade 
Critical Settings below for more information and to the 
OPTION:USE setting page20. 
The table below describes when the counter(s) will increment. 
Industrial or OIML: The Calibration Counter increments when 
trade critical settings are changed.  An example of the counter is 
C.00019. 
NTEP: Two counters display.   
The Calibration Counter increments when trade critical settings 
in the Calibration (CAL) menu are changed.  An example of the 
counter is C.00010. 
The Configuration Counter increments when other trade critical 
settings (ie. not in the CAL menu) are changed.  An example of 
the counter is F.00009. 
 
7.1.1.  Trade Critical Settings 
Trade critical settings can affect calibration and/or legal for trade 
performance. In this document the ⊗ symbol indicates the setting 
is trade critical.  Each time a trade critical setting is altered, the 
calibration counter will be incremented by one.   
7.2. rin-LINK 
For information on setting up using the rin-LINK cable refer to rin-
LINK page8.