Rinstrum -  R300 Series Digital Indicator Quick Start Manual Rev 1.5 
Page 20 
GROUP 
(GRP) 
 
ITEM 
(ITM) 
 
SELECT 
(SEL) 
 
EDIT 
(EDT) 
or   
Underline = Defaults 
OK 
 
 or   
BUILD 
DP⊗ 
Decimal Point Position  000000, 00000.0, 0000.00, 000.000, 00.0000, 0.00000  Accept 
 
CAP⊗ 
Maximum Capacity 
000100 to 999999  Default = 003000
 
<SEL> changes position, <EDT> changes digit. 
Accept 
 
RES⊗ 
Resolution (Count-By) 
1, 2, 5, 10, 20, 50, 100  Accept 
 
UNITS⊗ 
Weighed Units  none, g, kg, lb, t   Accept 
 
HI.RES⊗ 
Resolution x 10 Mode  OFF, ON  Accept 
 
CABLE⊗ 
4-Wire or 6-Wire  4, 6 Accept  
OPTION 
USE⊗ 
Scale Use  INDUST, OIML, NTEP  Accept 
  FILTER  Reading Average  
(time in seconds) 
none,  0.2,  0.5,  1.0,  2.0,  3.0,  4.0  Accept 
 
MOTION⊗ 
Motion Detection  OFF,  0.5-1.0,  1.0-1.0 
Default = 0.5 Graduations per Second 
Accept 
  INIT.Z  Initial Zero on Startup  OFF, ON  Accept 
 
Z.TRAC⊗ 
Zero Tracking 
Sensitivity 
OFF
, SLOW (0.5 grads/sec), FAST (10 grads/sec) 
Accept 
 
Z.RANGE⊗ 
Zero Operating Range  -2_ 2,  -1_ 3,  -20_20 
Default = -2_ 2 (–2% to +2%) 
Accept 
 
Z.BAND⊗ 
Zero 'Dead' Band 
<SEL> changes position, <EDT> changes digit. 
Default = 0
 
Accept