S00 to S31: Select a single unit with the matching address 00 to 31.
S96: De-select all units.
S97 & S98: All units are selected but none reply to commands. This mode is very useful for blanket
commands for an entire network of units.
S99: Select all units and all respond. S99 is useful when a single unit is connected to the network as it
is possible to select this unit regardless of its address setting.
Example
Command Answer Description
S01; Select unit 1
MSV?; 00400.0 CRLF Query current weight
S02; Select unit 2
MSV?; 00623.5 CRLF Query current weight
S96; De-select all units
21.3.47 STP: Stop Continuous Data
Stop continuous weight transmission started by MSV?,0; command.
General
No. of parameters 0
Save changes -
Increment Trade Counter -
Example
Command Answer Description
S01; Select unit 1
MSV?,0; 00400.0 CRLF
00400.1 CRLF
00400.2 CRLF
. . .
Start continuous data transmission
STP; Stop continuous data transmission
21.3.48 TAR: Tare
Force a tare operation, and read the current tare value.
General
No. of parameters 0 (command) / 1 (query)
Save changes At input
Increment Trade Counter No
This command is exactly the same as pressing the tare key on the front of the instrument except that the
indicator does not wait for no motion. If the current weight reading is not stable the indicator will return 1
CRLF and ignore the TAR command.
A TAR? query will return the currently applied tare value.
Parameter Description Range Default
1 Current tare value (query only) 000000 - 999999 -
Table of possible replies:
154 C500-600-2.0.3