01Series Operator’s Manual Alarm Points 18
NOTE: If you press and hold the POWER/MODE button for too long
(about five seconds), you will turn off the unit.
6. Use the AIR button to change the high alarm point setting. Pressing and
releasing the AIR button in quick succession increases the high alarm
point one number at a time. Pressing and holding the AIR button
increases the high alarm point at a faster rate.
NOTE: You can only cycle through the alarm points once before the 01
Series goes into its start-up sequence, which then places the unit
in Measuring Mode. If you want to cycle through the alarm points
again, press and hold the POWER/MODE button to turn off the 01
Series. Then begin with step 2 above to put the unit back into
Alarm Point Adjustment Mode.
7. When you are finished viewing or adjusting the alarm point settings,
press and release the POWER/MODE button until the ROM number of
the unit appears on the LCD. (The ROM is the component that contains
the software that runs the 01 Series.) The 01 Series will then go into its
start-up sequence followed by Measuring Mode.