1.4 Ambient requirements
All items and surfaces close to the machine must be made of non-flammable
materials, or have non-flammable coatings. The equipment must be oper-
ated at the ambient temperature from +5
to +40
C and relative humidity
not more than 45% at 40
C. The temperature decreasing related to RH in-
creasing, for example, 90% of RH at 20
C. Altitude above sea level should
not exceed 1000 m.
1.5 Power requirements
Power outlet MUST have grounding contact!
If supply cord damaged, it must be replaced
by manufacturer, service agent, or qualified
persons in order to avoid hazard!
Connections must be done by qualified electri-
cian ONLY!
Use IEC 60309 2P+PE 16 A plug:
Equipotential bonding wire (up to 10 mm
2
) shall be con-
nected to screw terminal marked with IEC 5021 sign:
6