Procedure for coding Accutrend
®
GC
Switch on the meter. When the function test is complete
the word CODE flashes. If the instrument has not been
coded before, the display shows
---
. If it has been
coded previously, the “old” three-digit code is displayed.
Take the code strip out of the wrapping and hold it
between the thumb and forefinger with the arrows
pointing away from you. Be careful not to touch the
black horizontal line.
With the flap
D closed, insert the code strip into the
slot
F in one smooth, brisk movement as far as it will
go, then withdraw the strip immediately. If the beeper is
switched on, successful coding is acknowledged by a
beep. The display shows a three-digit code number. The
code has been read in and stored.
If the error message
e-2
or
e-3
appears, coding was
unsuccessful (see Section 6.1). In this case, repeat the coding
as follows: open flap, close flap, insert code strip and then
withdraw again immediately.