3.3 Measurements on several people
If the Accutrend
®
GC is to be used in the medical sector for
measuring blood glucose and cholesterol in several different
persons, special measures must be observed. The blood must
be applied to the test strips outside the meter. This is very
important as otherwise the flap and test-strip guide may
become contaminated with blood residues and there is a risk
of infection.
In order to prevent infection (e.g. hepatitis or AIDS) it is
essential to observe the following points:
Only apply the blood outside the meter. Follow the
instructions in the section “Procedure for measurement
on several people” below.
Use a separate lancet and lancing device for each
person. We recommend using the Accu-Chek
®
Softclix
®
Pro lancing device and Accu-Chek
®
Softclix
®
Pro
Lancet, which were specially developed for this
purpose.
Disinfect the meter with 70% alcohol after each mea-
surement (see Section 5.2).