5.
76 77
5.4 Measurement and storage conditions
Please maintain the following conditions scrupulously every
time you perform a measurement. They are essential to ensure
precision and reliability of the measurements.
Temperature range for measurements
Accutrend
®
Glucose: +18°C to +32°C
Accutrend
®
Cholesterol: +18°C to +30°C
If a measurement is performed at a temperature outside the
ranges given above, the safety warning
TTT
flashes on the
display in alternation with the measured value.
The reading was obtained under temperature condi-
tions outside the admissible range. It must therefore be
interpreted with particular caution and may only be
used for orientation.
Allow the meter and test strips to adjust slowly to room tem-
perature. Never try to speed up the temperature adjustment by
placing on a radiator or in a refrigerator.
011553 AccuGC GB 19.02.2002 10:40 Uhr Seite 76