3.
42 43
Procedure for measurement on several people
Begin as described in Section 3.2.
Open the flap.
After lancing the finger remove the test strip from the
open meter and place on a piece of clean, absorbent
material, e.g. paper towel. Then apply the drop of blood
as described.
Immediately re-insert the test strip and close the flap.
Continue as described in Section 3.2.
Repeat the described procedure for the next person to be test-
ed, using a new piece of paper towel or similar absorbent
material.
If Accutrend
®
GC is used for performing measurements on sev-
eral people and the blood is always applied to the strip using
an application device (e.g. capillary pipettes) the sample may
also be applied with the test strip in the instrument (see Section
3.2). In this case a new application device must be used for
each measurement.