6.
86 87
The code strip is not suitable for the meter.
(e.g. a BM-Lactate code strip was inserted).
Use only Accutrend
®
Glucose or Accutrend
®
Cholesterol code strips.
A strong magnetic field is affecting the meter.
Move the meter to a different place
(see Section 5.4).
The optics are dirty or the meter is defective.
Clean the meter (see Section 5.2) and repeat the
measurement. If the problem persists, contact the
Roche Diagnostics service centre (see Section 6.3).