Software and Firmware Update
R&S
®
DDF205
60Getting Started 4073.0187.02 ─ 03
4.3.2 Firmware Update Using Update32 Tool
First steps
● Power off the R&S DDF205. It has to be switched on later in the update proce-
dure.
● Make sure the R&S DDF205 and the PC running the "Update32" tool are con-
nected either directly using a LAN cable or through the same network switch/
hub. It is NOT recommended that the PC and the R&S DDF205 be connected
across different routers when performing firmware update with "Update32"
because
special network packets required by the process may be discarded by
the router.
●
Run the "Update32" tool by clicking on Firmware Update Program. Allow
access if there are security warnings. Below is a screenshot of the Update32
application.
Selecting the configuration file
● To update a R&S DDF205, the correct update configuration file has to be
selected. Click on the following button to open the corresponding dialog:
●
Navigate to the folder that contains the configuration files under the "Look in:"
field.
Next, select the desired configuration file in the upper portion of the dialog
and click the "ACCEPT" button.
Receiver Firmware Update for DDF205 without front control panel