EasyManua.ls Logo

Rohde & Schwarz DDF205 - Rack Mounting; In-Vehicle Mounting; EMI Protective Measures; Connecting the R&S DDF205 to the Power Supply

Rohde & Schwarz DDF205
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparation for Use
R&S
®
DDF205
8Getting Started 4073.0187.02 ─ 03
1.2.2 Rack Mounting
Ambient temperature
The
R&S DDF205 should be used in an area where the ambient temperature
does not exceed –10 °C to +55 °C (DDF205 without front control panel) or
0 °C to 55 °C (DDF205 with front control panel), respectively. The
R&S DDF205 is fan-cooled and must be installed with sufficient space along
the sides to ensure a free flow of air. Make sure that there is sufficient space
for hot air to escape from the R&S DDF205. To ensure sufficient cooling do
not attach telescopic rails to the sides of the unit.
1.2.3 In-vehicle Mounting
For use in vehicles, the R&S DDF205 can be powered directly from the vehicle's
battery via the DC input.
1.2.4 EMI Protective Measures
In order to avoid electromagnetic interference (EMI), the R&S DDF205
may only be
operated when it is closed and all shielding covers are in place. Use only appropriate
shielded signal and control cables with proper termination.
1.2.5
Connecting the R&S DDF205 to the Power Supply
Connect the R&S DDF205 observing the following sections and instructions for use.
The R&S DDF205 is suitable only for DC operation.
1.2.5.1 Connecting to the Power Adapter
The R&S DDF205 is connected to the AC supply 100 V to 240 V via an external
AC/DC
power adapter and to the socket X1 (10 VDC to 32 VDC) on the rear panel.
Recommended connector: Neutrik® Speakon® NL4FX (see the figure below).
Setup