Common Measurement Settings
R&S
®
ESW
160User Manual 1177.6298.02 ─ 08
Generator Type
Selects the generator type and thus defines the generator setup file to use.
For an overview of supported generators, see "Overview of Supported Generators"
on page 154. For information on generator setup files, see "Generator Setup Files"
on page 155.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE on page 493
Interface
Type of interface connection used.
For details on which signal generators support which interfaces, see the documenta-
tion of the corresponding signal generator.
GPIB
TCP/IP
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface on page 492
TTL Handshake
If available for the specified generator type, this option activates TTL synchronization
via handshake.
Using the TTL interface allows for considerably higher measurement rates, because
the frequency stepping of the R&S ESW is directly coupled with the frequency stepping
of the generator.
For more information on TTL synchronization, see "TTL synchronization" on page 156.
For an overview of which generators support TTL synchronization see "Overview of
Supported Generators" on page 154.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK on page 492
GPIB Address/TCPIP Address / Computer Name
For LAN connections: TCP/IP address of the signal generator
For GPIB connections: GPIB address of the signal generator.
Remote command:
SYSTem:COMMunicate:GPIB:RDEVice:GENerator<gen>:ADDRess on page 492
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<gen>:ADDRess
on page 493
Reference
Selects the internal R&S ESW or an external frequency reference to synchronize the
R&S ESW with the generator (default: internal).
Remote command:
SOURce<si>:EXTernal<gen>:ROSCillator[:SOURce] on page 490
Edit Generator Setup File
Displays the setup file for the currently selected Generator Type in read-only mode in
an editor.