Remote Control
R&S
®
FPC
250User Manual 1178.4130.02 ─ 07
[SENSe:]FREQuency:STARt
[SENSe:]FREQuency:STOP
Parameters:
<OperationMode> CW
Select the time domain (span = 0)
FIXed
Select the time domain (span = 0)
SWEep
Select the frequency domain (span > 0).
*RST: SWEep
Example: //Activate frequency domain measurements
FREQ:MODE SWE
Manual operation: See "Frequency Mode" on page 368
[SENSe:]FREQuency:OFFSet
<FrequencyOffset>
This command defines a frequency offset.
Parameters:
<FrequencyOffset> <numeric value>
Range: depends on the start frequency to 100 GHz
*RST: 0 Hz
Default unit: Hz
Example: //Define a frequency offset
FREQ:OFFS 100MHZ
Manual operation: See "Frequency Offset" on page 368
See "Frequency Offset" on page 393
[SENSe:]FREQuency:SPAN <Span>
This command defines the frequency span.
If you define a span of 0 Hz, the R&S FPC starts a measurement in the time domain.
Parameters:
<Span> <numeric value>
Refer to the datasheet for details on the frequency range of your
R&S FPC.
*RST: fmax with fmax = maximum frequency
Default unit: Hz
Example: //Define frequency span
FREQ:SPAN 10MHZ
Manual operation: See "Manual Span" on page 369
See "Zero Span" on page 369
See "Manual Span" on page 399
Remote Commands of the Spectrum Application

Table of Contents

Other manuals for Rohde & Schwarz FPC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz FPC Series and is the answer not in the manual?

Rohde & Schwarz FPC Series Specifications

General IconGeneral
BrandRohde & Schwarz
ModelFPC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals