EasyManua.ls Logo

Rohde & Schwarz FSH13 - Page 124

Rohde & Schwarz FSH13
380 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSH Spectrum Analyzer Mode
Configuring Spectrum Measurements
Operating Manual 1173.6275.12 - 30 109
The R&S FSH opens a softkey submenu to control the gate settings. At the same time,
the R&S FSH switches into time domain, as indicated in the display.
Press the "Manual SWP Time" softkey and set the sweep time in a way that the
portion of interest of the signal is visible on the screen.
Press the "Gate Delay" softkey.
The R&S FSH opens an input field to define the trigger delay.
Enter the delay time you need.
The measurement now starts after the delay time has passed.
Press the "Gate Length" softkey.
The R&S FSH opens an input field to define the gate length.
Enter the length of the gate.
The R&S FSH now measures over the period of the gate length. After the gate has
closed, the R&S FSH waits with the measurement until the next gate signal
happens.
The delay time and gate length are represented by vertical red lines in the diagram
area.
After setting the gate delay and gate length, exit the gate settings menu with the
"Exit" softkey.
The R&S FSH returns to the frequency domain provided it was active before
setting up the gated trigger. The original span is restored. The R&S FSH is ready
to perform measurements with an accurately set gate.
3.2.4.4 Selecting the Sweep Type
The sweep type defines the way how the R&S FSH captures and processes the
data.
Press the SETUP key.
Press the "User Preferences" softkey.
Select the "Sweep / FFT" menu item.
The R&S FSH opens a dropdown menu to select the sweep type.
Which sweep type is appropriate for the current measurement depends on the
span, RBW, VBW and sweep time settings.
- Auto
The R&S FSH automatically selects the fastest sweep type for the current
measurement configuration.
- Always Sweep
The R&S FSH uses the local oscillator to provide the spectrum quasi analog
from the start to the stop frequency.