R&S FSH Spectrum Analyzer Mode
Configuring Spectrum Measurements
Operating Manual 1173.6275.12 - 30 120
3.2.6.7 Using Marker Search Limits
The R&S FSH allows you to use only a limited section of the trace for the "Set to
Peak", "Set to Next Peak" and "Minimum" functions.
Press the MKR key.
Press the "Search Limits" softkey.
Select the "Search Limits On/Off" menu item.
Confirm the selection with the ENTER key.
The R&S FSH activates the marker search limits.
An [X] indicates an active search limit. Two vertical red lines show the lower and upper
limits in the diagram.
By default, the search limit range is over the whole span.
Press the "Search Limits" softkey
Select the "Lower Limit" menu item.
Confirm the selection with the ENTER key.
The R&S FSH opens an input field to define the lower limit of the search range.
Enter the lower limit.
Confirm the entry with one of the unit keys.
If the span is wide enough, the R&S FSH displays a red vertical line to indicate the
lower limit.
Define the upper search limit the same way.
Deactivating marker search limits
Press the "Search Limits" softkey.
Select the "Search Limits On/Off" menu item.
Confirm the selection.
The "Search Limits" softkey turns grey again and in the "Search Limits" menu, the
[X] is no longer displayed.