EasyManua.ls Logo

Rohde & Schwarz FSH13 - Carrier Frequency Error

Rohde & Schwarz FSH13
380 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S FSH Digital Modulation Analyzer
Measurements on CDMA2000 Signals
Operating Manual 1173.6275.12 - 30 276
PN Found
The PN offset found during auto PN offset operation.
Tau
According to the CDMA2000 standard, Tau shows the timing error of the signal. The
maximum offset is specified at 10 µs.
Active Channels
Number of currently active channels.
9.5.1.3 Channel Results
Channel results contain various results specific to one or more channels. The table
contains
the absolute channel power in dBm
the channel power in relation to the total signal power in dB
the channel power in relation to the pilot channel in dB
for the pilot channel (PICH) and the synchronization channel (SYNC).
The pilot channel always occupies code 0 and the synchronization channel always
occupies code number 32.