Remote Commands
R&S
®
FSW
791User Manual 1173.9411.02 ─ 19
Measuring the Mean Power
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:AVERage:RESult?.......................791
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:PHOLd:RESult?.......................... 791
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult?......................................791
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:AVERage:RESult?
This command queries the average mean time domain power. The query is only possi-
ble if averaging has been activated previously using CALCulate<n>:MARKer<m>:
FUNCtion:SUMMary:AVERage on page 788.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Return values:
<MeanPower> Mean power of the signal during the measurement time.
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:PHOLd:RESult?
This command queries the maximum mean time domain power. The query is only pos-
sible if the peak hold function has been activated previously using CALCulate<n>:
MARKer<m>:FUNCtion:SUMMary:PHOLd.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Return values:
<MeanPower> Mean power of the signal during the measurement time.
Usage: Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN:RESult?
This command queries the mean time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Return values:
<MeanPower> Mean power of the signal during the measurement time.
Usage: Query only
Manual operation: See "Results" on page 267
Configuring and Performing Measurements