Analysis
R&S
®
FSWP
70User Manual 1177.5856.02 ─ 07
"Full"
Displays the Spectrogram in a subwindow in the full size of the origi-
nal result display.
"Off"
Closes the Spectrogram subwindow.
Remote command:
CALCulate<n>:SPECtrogram:LAYout on page 147
Trace
Selects the diagram trace on which the spectrogram is based.
Remote command:
CALCulate<n>:SGRam:TRACe on page 147
6.2 Marker Settings
Access
●
"Overview" > "Analysis" > "Marker"
●
"Overview" > "Analysis" > "Marker Function"
For more information, refer to the R&S FSWP user manual.
Branch for Peaksearch .................................................................................................70
Branch for Peaksearch
Defines which data is used for marker search functions in I/Q data.
This function is only available for the display configuration "Real/Imag (I/Q)" (see "
Real/Imag (I/Q) " on page 16).
Note: The search settings apply to all markers, not only the currently selected one.
"Real"
Marker search functions are performed on the real trace of the I/Q
measurement.
"Imag"
Marker search functions are performed on the imaginary trace of the
I/Q measurement.
"Magnitude"
Marker search functions are performed on the magnitude of the I and
Q data.
Remote command:
CALCulate<n>:MARKer<m>:SEARch on page 146
6.3 Display Lines and Limit Lines
Access (limit lines): "Overview" > "Analysis" > "Lines" > "Limit Lines"
Access (display lines): "Overview" > "Analysis" > "Lines" > "Display Lines"
Display Lines and Limit Lines