Two-port measurements with the tracking generator  R&S FSH 
1145.5973.12 4.100  E-11 
To shift in the y direction: 
 Press the MOVE Y softkey. 
 Set a value from  -50% to +50% by using the rotary 
knob or cursor keys or by entering a numeric value. 
 Confirm the selection with the ENTER key or the 
MOVE Y softkey. 
 
To enlarge an area: 
 Press the ZOOM IN softkey. 
The selected window area will be enlarged by the 
zoom factor that has been set. You can fine-adjust the 
zoom window by using the MOVE X and MOVE Y as 
described. 
To deactivate enlargement: 
 
 Press the ZOOM OUT key. 
 
Spectrum measurements with the VSWR Bridge R&S FSH-Z3 or R&S FSH-Z2 
connected 
To localize interfering signals that can affect the reflection measurement on an antenna, for example, it 
is useful to switch to the spectrum display. To eliminate having to dismount the VSWR bridge each 
time, a typical value for the insertion loss of the VSWR bridge is incorporated in the measurement. This 
corrective step requires that the control cable for detecting the bridge be connected to the R&S FSH. 
Since the value is merely a typical frequency-independent correction value, an additional level 
measurement uncertainty of max. 2 dB must be anticipated. 
 
Operation: 
 Press the MEAS key. 
 Select the MEASURE softkey. 
 Select ANALYZER from the menu by using the 
rotary knob or the cursor keys. 
As soon as the control cable has been connected to 
the R&S FSH, the VSWR Bridge R&S FSH-Z3 or R&S 
FSH-Z2 is automatically detected and indicated both 
on the display and the status menu. Automatic 
detection requires that this feature be activated in the 
SETUP menu (default setting).