Measurements on CW Signals R&S FSH
1145.5973.12 2.12 E-11
Calibrate the R&S FSH for the return loss measurement.
The following example shows a scalar measurement of return loss. If the option R&S FSH-K2 is
installed, measurement must first be switched to scalar.
! Press the MEAS key.
! Press the MEAS MODE softkey.
! Using the rotary knob or cursor keys, select SCALAR.
! Confirm with the ENTER key or the MEAS MODE softkey.
! In the main menu for the tracking generator, press
the REFLECT CAL softkey.
The R&S FSH prompts you to leave the measurement
port open.
! Leave the test port of the R&S FSH-Z2 open.
! Using the CONTINUE softkey, start the OPEN cali-
bration.
While calibration is in progress, the R&S FSH outputs
the message "Calibrating OPEN, please wait... ".
When OPEN calibration is over, the R&S FSH prompts you to perform SHORT calibration.
! Connect a short to the test port of the R&S FSH-Z2.
! Using CONTINUE, start the SHORT calibration.
While calibration is in progress, the R&S FSH outputs
the message "Calibrating SHORT, please wait... ".
When calibration is over, the R&S FSH outputs the
message "Reflect. calibrated" for 3 seconds.
Reflection is displayed in the upper right-hand corner
of the measurement diagram to indicate that the
R&S FSH is calibrated for reflection measurements.
! Connect the DUT to the measurement port of the
R&S FSH-Z2 or R&S FSH-Z3.
The R&S FSH displays the return loss of the DUT.