Remote control commands
R&S
®
RTM3000
445User Manual 1335.9090.02 ─ 09
PROBe<m>:ID:SRNumber?
Returns the serial number of the probe.
Suffix: 
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values: 
<SerialNumber> Serial number in a string
Usage:  Query only
PROBe<m>:ID:SWVersion?
Returns the version of the probe firmware.
Suffix: 
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values: 
<SoftwareVersion> Version number in a string.
Usage:  Query only
PROBe<m>:SETup:TYPE?
Returns the type of the probe: not detected, active or passive probe.
Suffix: 
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values: 
<Type> NONE | ACTive | PASSive
Usage:  Query only
PROBe<m>:SETup:NAME?
Returns the name of the probe.
Suffix: 
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values: 
<Name> String parameter
Example: 
PROB:SET:NAME?
<-- "RT-ZS20"
Usage:  Query only
Waveform setup