PROBe<m>:SETup:BANDwidth?
Returns the bandwidth of the probe.
Suffix: 
<m>
.
Selects the input channel. The number of channels depends on
the instrument.
Return values: 
<Bandwidth> Range:  10e5  to  20e8
Increment:  10
Default unit: Hz
Usage:  Query only
PROBe<m>:SETup:ATTenuation[:AUTO]?
Returns the attenuation of an automatically detected probe.
Suffix: 
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Return values: 
<ProbeAttenuation> Range:  0.0001  to  10000
Usage:  Query only
PROBe<m>:SETup:GAIN[:AUTO]?
Returns the gain of an automatically detected probe.
Suffix: 
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Return values: 
<ProbeGain> Range:  0.0001  to  10000
Usage:  Query only
PROBe<m>:SETup:CAPacitance?
Returns the input capacity of the probe.