Preparing for use
R&S
®
SMA100B
24Getting Started 1419.8620.02 ─ 10
When connected, the R&S SMA100B detects the keyboard automatically. A
detected keyboard has the default layout English – US.
To connect a mouse
Connect the mouse to any of the USB connectors.
When connected, the R&S SMA100B detects the mouse automatically.
To connect power sensors
You can connect power sensors of the R&S NRP families to any of the USB con-
nectors on the front panel or rear panel.
See chapter "Using power sensors" in the user manual.
4.9 Connecting to RF coaxial connectors
Here, you find information on how to prepare and to connect to RF coaxial con-
nectors of the R&S SMA100B. Use these RF connectors, for example, for output
of the RF signal or for input of an external reference signal.
To prepare for connecting
1. NOTICE! Damaged or not clean connections can lead to RF insertion loss
and mismatch, and even premature wear of the connectors.
Before connecting to the port, inspect the RF connector visually. Check that it
is clean, undamaged and mechanically compatible.
2. NOTICE! DC voltage at the RF connector can damage the instrument. Never
apply DC voltage to the RF input connectors.
Make sure that the values are within the DC limits given in the data sheet.
3. If your test setup has a DC component at the RF input, insert a DC blocker.
4. Use a high-quality RF cable that matches the RF connector type.
See "Cable selection and electromagnetic interference (EMI)" on page 21.
5. You can connect to two kinds of connectors:
"To connect to screwable connectors" on page 25
"To connect to pluggable connectors" on page 25
Connecting to RF coaxial connectors