Command Reference
R&S
®
ZNB/ZNBT
1098User Manual 1173.9163.02 ─ 54
<TraceFile> Mandatory string parameter containing the path and file name of
the file. If the path is omitted, the current directory is used (see
MMEMory:CDIRectory).
Usage: Setting only
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Export Test Results" on page 653
MMEMory:STORe:EYE:MEASurements <TraceName>, <TraceFile>[,
<DecSeparator>, <FieldSeparator>]
Allows to export eye diagram results to an ASCII file (csv), with the possibility to spec-
ify a decimal separator (comma or point) and a field separator (semicolon, comma, tab
or space).
See also CALCulate<Chn>:EYE:MEASurement:DATA? on page 815.
Note that the decimal separator and field separator must be different: if both are set to
comma, actually a semicolon will be used as field separator.
Setting parameters:
<TraceName> The name of the eye diagram whose results shall be exported
<TraceFile> Mandatory string parameter containing the path and file name of
the file. If the path is omitted, the current directory is used (see
MMEMory:CDIRectory).
<DecSeparator> POINt | COMMa
Decimal separator
*RST: COMMa
<FieldSeparator> SEMicolon | COMMa | TABulator | SPACe
Field separator
*RST: SEMicolon
Usage: Setting only
Options: R&S ZNB/ZNBT-K20
Manual operation: See "Export Measurements..." on page 640
MMEMory:STORe:LIMit <TraceName>, <LimLineFile>
Saves the limit lines associated to a specified trace to a limit line file. Limit lines are
created using the CALCulate<Chn>:LIMit... commands.
Setting parameters:
<TraceName> Name of an existing trace in the active recall set (string parame-
ter) for which a limit line definition exists.
SCPI Command Reference