Command Reference
R&S
®
ZNB/ZNBT
920User Manual 1173.9163.02 ─ 54
Parameters:
<EvalRange> Number of the evaluation range.
Range: 1 to 10. In addition, 0 denotes the (non-configura-
ble) "Full Span" evaluation range.
*RST: 0
Example:
*RST; :CALC:STAT:DOM:USER?
Query the default evaluation range. The response is zero, i.e.
the evaluation range is equal to the complete sweep range
CALC:STAT:DOM:USER 1
CALC:STAT:DOM:USER:STARt 1GHZ; STOP 2GHZ; SHOW
ON
Select evaluation range no. 1 and define the evaluation range
between 1 GHz and 2 GHz. Display the range limit lines.
Manual operation: See "Evaluation Range" on page 331
CALCulate<Chn>:STATistics:DOMain:USER:SHOW <Boolean>
Displays or hides range limit lines for the evaluation range selected via
CALCulate<Chn>:STATistics:DOMain:USER.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - range limit lines on or off.
*RST: OFF
Example: See CALCulate<Chn>:STATistics:DOMain:USER
Manual operation: See "Range Limit Lines On" on page 331
CALCulate<Chn>:STATistics:DOMain:USER:STARt <Start>
CALCulate<Chn>:STATistics:DOMain:USER:STOP <Stop>
These commands define the start and stop values of the evaluation range selected via
CALCulate<Chn>:STATistics:DOMain:USER.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Stop> Start or stop value of the evaluation range.
Default unit: NN
Example: See CALCulate<Chn>:STATistics:DOMain:USER
Manual operation: See "Evaluation Range" on page 331
SCPI Command Reference