EasyManua.ls Logo

Rohde & Schwarz ZNB - Calculate<Ch>:Transform:vnetworks:gloop:group

Rohde & Schwarz ZNB
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
970User Manual 1173.9163.02 ─ 54
Example:
*RST; :CALC:TRAN:VNET:GLO:EMB:PAR:R? SC; R? SL
Query the default resistances for ground loop embedding. The
response is 10000000; 0.
CALC:TRAN:VNET:GLO:EMB:PAR:R SC, 2.2E+3
Increase the resistance for the Shunt C model to 2.2 kΩ.
Manual operation: See "Network" on page 610
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
TNDefinition <CircuitModel>
Selects the circuit model for ground loop embedding.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<CircuitModel> FIMPort | SL | SC | SG
Possible circuit models (character data); see Circuit models for
ground loop port embedding/deembedding.
*RST: FIMPort
Example: See CALCulate<Ch>:TRANsform:VNETworks:GLOop:
EMBedding<group>[:STATe] on page 966
Manual operation: See "Network" on page 610
CALCulate<Ch>:TRANsform:VNETworks:GLOop:GROup <Boolean>
If set to ON (1), each port group (defined using SOURce<Ch>:GROup<Grp>:PPORts)
has its own ground loop embedding and deembedding network. Otherwise the same
embedding network and deembedding network is used for all active ports.
See the
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:...
and CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:...
commands.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> Range: ON (1) | OFF (0)
*RST: OFF (0)
Manual operation: See "Ground Loop per Port Group" on page 611