Command Reference
R&S
®
ZNB/ZNBT
878User Manual 1173.9163.02 ─ 54
<Mk> Marker number
Parameters:
<Value> Minimum peak excursion
The unit is derived from the active trace format and cannot be
modified. Changing the trace format resets the excursion to a
format-specific default value.
Manual operation: See "Excursion Settings" on page 371
CALCulate<Chn>:MARKer<Mk>:EXCursion:STATe <Boolean>
Activates or deactivates the marker-specific minimum excursion for peak searches.
Use CALCulate<Chn>:MARKer<Mk>:EXCursion to set the minimum peak excur-
sion.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number
Parameters:
<Active> Boolean
Manual operation: See "Excursion Settings" on page 371
CALCulate<Chn>:MARKer<Mk>:FORMat <OutFormat>
Sets/queries the output format for the (complex) value of the related marker.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number.
Parameters:
<OutFormat> MLINear | MLOGarithmic | PHASe | POLar | GDELay | REAL |
IMAGinary | SWR | LINPhase | LOGPhase | IMPedance |
ADMittance | DEFault | COMPlex | MDB | MLPHase |
MDPHase | MIMPedance | PIMPedance | PADMittance |
MADMittance | MPIMpedance | MPADmittance | INDX
DEFault means that the marker is formatted according to the
related trace's default marker format (see CALCulate<Chn>:
MARKer:DEFault:FORMat), which may not be set to DEFault
(= trace format).
For the other marker formats, see CALCulate<Chn>:MARKer:
DEFault:FORMat and the description in "Marker Format"
on page 98.
*RST:
DEFault
Manual operation: See "Marker Format" on page 365