709User Manual 1173.9163.02 ─ 54
Remote command:
DIAGnostic:DUMP:SIZE
Touchstone Export Options
Configures whitespace insertion during Touchstone file export.
The default export format is explained in Chapter 5.4.2.1, "Touchstone Files",
on page 143:
logical columns are vertically aligned using spaces
positive and negative numbers are vertically aligned by prefixing positive numbers
with blanks
the frequencies are horizontally separated from the corresponding S matrices
using leading spaces
the content parts (header, S matrices for different frequencies) are separated by
blank lines
Use TAB (instead of blanks) ← Touchstone Export Options
If checked, columns are separated by tabs rather than spaces.
Remote command:
MMEMory:STORe:TRACe:OPTion:TABS
Trim Leading Whitespace ← Touchstone Export Options
If checked, whitespace at the beginning of each line is removed.
Remote command:
MMEMory:STORe:TRACe:OPTion:TRIM
Positive Number Prefix ← Touchstone Export Options
Positive numbers can either be prefixed by blanks, by plus signs or not at all.
Remote command:
MMEMory:STORe:TRACe:OPTion:PLUS
Skip Separator Lines ← Touchstone Export Options
If checked, the content parts are no longer separated by blank lines.
Remote command:
MMEMory:STORe:TRACe:OPTion:SSEParator
TTL Pass Default Values
Defines the default values of TTL1 Pass / TTL2 Pass for new traces.
Remote command:
SYSTem:TTLout<Pt>:STATus[:STATe]