Command Reference
R&S
®
ZNB/ZNBT
1201User Manual 1173.9163.02 ─ 54
<SourceNumber> Number of the port for the internal source
Range: 1 to port number of the analyzer/number of external
generators
*RST: n/a (for all parameters)
Example:
*RST; FREQ:IMOD:LTON PORT,1
Select port 1 as a source port for the lower tone.
FREQ:IMOD:UTON GEN,1
Select an external generator as a source for the upper tone
(must be configured previously).
SOUR:POWer:CORR:IMOD:PORT 'COMBINER'
Select Extenal Device as a source port for the two tone signal.
FREQ:IMOD:REC 2
Select port 2 as the receiver port for the intermodulation mea-
surement.
FREQ:IMOD:TDIS 2E6
Select a tone distance of 2 MHz.
FREQ:IMOD:ORD3 ON; :ORD5 ON; :ORD7 OFF; :ORD9
OFF
Enable the measurement of the intermodulation products up to
the 5th order.
...
(Perform intermodulation measurement, evaluate results.)
FREQ:IMOD:CONV OFF
Disable intermodulation measurement, switch back to normal
(non frequency-converting) mode.
Manual operation:
See "Lower Tone" on page 554
[SENSe<Ch>:]FREQuency:IMODulation:ORDer<Im>[:STATe] <Boolean>
Enables or disables the measurement of the intermodulation products of order <IM
order>.
Suffix:
<Ch>
.
Channel number
<Im> Order of IM products
Parameters:
<Boolean> ON | OFF – enables or disables measurement
*RST: ON (for <Im> = 3), OFF (for <Im> > 3)
Example: See [SENSe<Ch>:]FREQuency:IMODulation:LTONe
Manual operation: See "Prepare Measurement of IM Order" on page 555
[SENSe<Ch>:]FREQuency:IMODulation:RECeiver <PortNumber>
Selects the receiving port for the intermodulation measurement.
SCPI Command Reference