Command Reference
R&S
®
ZNB/ZNBT
1231User Manual 1173.9163.02 ─ 54
<LO> POSitive | NEGative | AUTO
Position of the local oscillator frequency LO relative to the RF
frequency. In remote control this parameter must be set when a
sweep segment is created.
See [SENSe<Ch>:]FREQuency:SBANd.
<Selectivity> NORMal | MEDium | HIGH
Selectivity of the IF filter. See [SENSe<Ch>:]SEGMent<Seg>:
BWIDth[:RESolution]:SELect.
<Mode> AUTO
Ignored.
<FreqSweepMode> STEPped | ANALog
"Stepped"mode (STEPped, default) or "Swept" mode
(=ANALog).
Example:
SEGM:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.5S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz.
SEGM2:ADD
Create a second sweep segment. The frequency range of the
second segment will be between 1.5 MHz and the maximum fre-
quency of the analyzer.
Manual operation: See "Add / Insert / Delete / Delete All" on page 403
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect <TimeRef>
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the
command [SENSe<Ch>:]SEGMent<Seg>:INSert, is entered as a segment sweep
time ("Segment Time") or as a measurement delay ("Meas Delay").
Suffix:
<Ch>
.
Channel number
<Seg> Sweep segment number
Parameters:
<TimeRef> SWTime | DWELl
SWTime - use segment sweep time.
DWELl - use measurement delay.
*RST: SWTime