Command Reference
R&S
®
ZNB/ZNBT
1315User Manual 1173.9163.02 ─ 54
Example:
SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter 1',
'NRP-Z55', 'usb', '100045'
SYST:COMM:RDEV:PMET2:DEF 'USB Power Meter 2',
'NRP-Z55', 'usb', '100046'
Configure two R&S NRP power meters as external power meter
no. 1 and 2, assigning names and serial numbers.
SYST:COMM:RDEV:PMET:CAT?
Query the power meter numbers. The response is '1, 2'.
SYST:COMM:RDEV:PMET:COUN?
Query the number of configured power meters. The response is
2.
Usage: Query only
Manual operation: See "Configured Devices" on page 722
SYSTem:COMMunicate:RDEVice:PMETer:CONFigure:AUTO[:STATe]
<Boolean>
Enables or disables auto-configuration of R&S NRPxxS/SN power sensors. If the func-
tion is enabled, the analyzer automatically configures all power sensors detected at
any of the USB ports as Pmtr 1, Pmtr 2, ...
Parameters:
<Boolean> ON | OFF – enable or disable auto-configuration.
Example: See SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:
DEFine
Manual operation: See "Auto Config" on page 724
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
Queries the number of configured external power meters. The result is an integer num-
ber of power meters.
Example:
See SYSTem:COMMunicate:RDEVice:PMETer:CATalog?
Usage: Query only
Manual operation: See "Configured Devices" on page 722
SYSTem:COMMunicate:RDEVice:PMETer:DELete
Clears the configuration table for external power meters.
Example:
See SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:
DEFine
Usage: Event
Manual operation: See "Configured Devices" on page 722
SCPI Command Reference