Command Reference
R&S
®
ZNB/ZNBT
1342User Manual 1173.9163.02 ─ 54
SYSTem:SOUNd:ALARm[:STATe] <Boolean>
SYSTem:SOUNd:STATus[:STATe] <Boolean>
These commands switch alarm or status sounds on or off.
Parameters:
<Boolean> ON | OFF
*RST: ON
Example:
SYST:SOUN:ALAR OFF; STAT OFF
Switch alarm and status sounds off.
Manual operation: See "Sounds" on page 704
SYSTem:TIME <Hours>, <Minutes>, <Seconds>
The command queries or defines the instrument's current time setting.
The setting command requires administrator rights; refer to Chapter 11.1.1, "Windows
Operating System", on page 1425.
Parameters:
<Hours> Range: 0...23
<Minutes> Range: 0...59
<Seconds> Range: 0...59
Example:
SYST:TIME?
Response: 12,0,0 - it is precisely 12 pm.
SYSTem:TRESet[:STATe] <Boolean>
Defines the behavior of the INITiate<Ch>[:IMMediate][:DUMMy] and
INITiate[:IMMediate]:ALL.
Parameters:
<Boolean> OFF (0)
Previous trace data are preserved.
ON (1)
Previous trace data are deleted.
This is a global setting that is not affected by an instrument
reset. Factory default is OFF (0).
Manual operation: See "Restart: Set all Traces to 0" on page 705
SYSTem:TSLock <Type>
Locks the touchscreen functionality of the R&S ZNB/ZNBT, e.g. in order to prevent
inadvertent entries during remote control.