EasyManua.ls Logo

Rohde & Schwarz ZNB - Page 259

Rohde & Schwarz ZNB
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GUI Reference
R&S
®
ZNB/ZNBT
259User Manual 1173.9163.02 ─ 54
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:SDEFine:SENDed
S<out><in>
Selects one of the four elements of the standard 2-port S-parameters as a measured
quantity for the active trace.
Remote command:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
All S-Params
With the active logical ports distributed to m DUTs with n
1
, ..., n
m
ports, respectively, the
analyzer creates n
1
2
+ ... + n
m
2
diagrams and displays the full set of S-Parameters, one
in each diagram. The diagrams are arranged as an n x n grid.
If "All S-Params" would result in more than 100 diagrams being created, a warning dia-
log is opened that lets you cancel the diagram creation.
The reflection coefficients S
ii
appear in the diagrams on the main diagonal, the trans-
mission coefficients S
ij
(i j) occupy the other diagrams. By default, reflection coeffi-
cients are displayed in Smith diagrams; transmission coefficients in Cartesian dia-
grams "dB Mag" scale.
See also Format Softtool.
Remote command:
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:DEFine:SGRoup
S-Param Wizard...
Opens a dialog that guides you through the setup of a standard multi-port S-parameter
measurement in a frequency sweep. See Chapter 6.2.1.2, "S-Parameter Wizard",
on page 260.
Dialog and opener button are not available for R&S ZNBT, or if a switch matrix is con-
figured.
Balanced Ports...
Opens a dialog that allows you to enable/disable physical ports and to define logical
ports (balanced or unbalanced) in the active channel. See Chapter 6.2.1.3, "Balanced
Ports Dialog", on page 262.
Note: If either multiple DUTs are configured (see Chapter 6.12.2.3, "Define Parallel
Measurement Dialog", on page 535) or the "Fixture Simulator" is disabled for the rela-
Meas Softtool

Table of Contents

Other manuals for Rohde & Schwarz ZNB

Related product manuals