GUI Reference
R&S
®
ZNB/ZNBT
442User Manual 1173.9163.02 ─ 54
The layout of the dialog panel depends on the number of test ports available.
Figure 6-22: Define Calibration dialog (manual, > 4 test ports)
Ports
Selects the test ports to be calibrated.
Note: Calibration and port de-/activation.
The way the analyzer fimware activates/deactivates ports after a successfull calibration
(system error correction or power calibration) has slightly changed:
In all FW releases, calibrated ports that were previously disabled, are automatically
enabled as single-ended logical ports.
Starting with FW version 2.10, uncalibrated ports were disabled in the related
channel(s). If only one of the physical ports forming a balanced port was calibrated,
Cal Softtool