716User Manual 1173.9163.02 ─ 54
Password
Enter a password here to activate the required service level.
Remote command:
SYSTem:PASSword[:CENable]
Service Function
Identifier of the service function in "dotted textual" (example: sw.common.mem-
ory_usage) or "dotted decimal" (example: 0.1.18.0) representation.
Remote command:
DIAGnostic:SERVice:SFUNction
6.17.2 Freq. Ref. Tab
Selects a reference signal for synchronization between the R&S ZNB/ZNBT and exter-
nal devices. A common reference frequency is advisable to ensure frequency accuracy
and frequency stability in the test setup.
State
Indicates the state of the internal phase locked loop: If the frequencies are properly
synchronized, the state is "locked".
Remote command:
n/a
Internal/External
Selects the internal or an external reference clock signal for synchronization.
Internal: The analyzer synchronizes to its internal 10 MHz reference clock.