GUI Reference
R&S
®
ZNB/ZNBT
741User Manual 1173.9163.02 ─ 54
Open the Info Dialog for full information about the instrument.
6.19 Additional Function Keys
The SYSTEM hardkey panel also provides the Windows®, [HELP] and [PRESET] keys
that do not open a softtool but perform an immediate action.
The DATA ENTRY panel provides hardkeys for undo and redo.
SYSTEM – [PRESET]
Performs a preset of all instrument settings or of the active recall set, depending on the
settings in the Presets Tab of the "System Config" dialog.
A preset can be a factory preset or a user-defined preset. It does not affect global set-
tings and resources (e.g. "System Config" settings, calibration kits and calibration pool
etc.).
Tip: If you perform a [PRESET] by mistake, use [UNDO] to restore your previous
instrument settings.
Remote command:
*RST
SYSTem:PRESet:SCOPe
SYSTem:PRESet:USER:NAME
SYSTem:PRESet:USER[:STATe]
SYSTem:PRESet[:DUMMy]
DATA ENTRY – [UNDO] / [REDO]
[UNDO] reverses the last action, [REDO] reverses a preceding [UNDO].