Command Reference
R&S
®
ZNB/ZNBT
866User Manual 1173.9163.02 ─ 54
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
If you want to use a custom formula for a segment, use
CALCulate<Chn>:LIMit:CONTrol[:DATA] to create it.
Parameters:
<State> OFF (0)
The line segment is a straight line, connecting the endpoints
specified in the segment definition.
ON (1)
The line segment is defined by a custom formula. Use
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula to spec-
ify the formula.
*RST: OFF
Example: See CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula
on page 865
Manual operation: See "Linear/Formula" on page 351
CALCulate<Chn>:LIMit:SEGMent<Seg>:INTerpol <LimLineInterpolation>
"Interpolation" determines whether the limit line segment is interpolated linearly or log-
arithmically. The latter may be more suitable for logarithmic sweeps.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number
Parameters:
<LimLineInterpolation>LOGarithmic | LINear
Manual operation: See "Segment List" on page 348
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STARt <FreqPowTime>
CALCulate<Chn>:LIMit:SEGMent<Seg>:STIMulus:STOP <StimVal>
These commands change the start and stop stimulus values (i.e. the smallest and the
largest stimulus values) of a limit line segment. A segment must be created first to
enable the commands (e.g CALCulate<Chn>:LIMit:DATA). See also Chap-
ter 5.4.1.1, "Rules for Limit Line Definition", on page 135.
Tip: To define the stimulus values of several limit line segments with a single com-
mand, use CALCulate<Chn>:LIMit:CONTrol[:DATA].
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Seg> Segment number