EasyManua.ls Logo

Rohde & Schwarz ZNB - Page 941

Rohde & Schwarz ZNB
1498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command Reference
R&S
®
ZNB/ZNBT
941User Manual 1173.9163.02 ─ 54
Example:
*RST; :CALCulate1:TRANsform:DTFault:STATe ON
makes the active trace of channel 1 a DtF trace.
SENSe1:SWEep:POINts?
returns the default number of sweep points (201).
:CALCulate1:TRANsform:DTFault:STARt 0; STOP 100
widens the DtF distance window to 100m.
CALCulate1:TRANsform:DTFault:POINts
starts the "Auto Number of Points calculation". Now
SENSe1:SWEep:POINts?
returns an increased number of sweep points.
Usage: Event
Options: R&S ZNB/ZNBT-K3
Manual operation: See "Auto Number of Points" on page 678
CALCulate<Chn>:TRANsform:DTFault:STARt <Start>
CALCulate<Chn>:TRANsform:DTFault:STOP <Stop>
Defines the window of the Distance to Fault measurement using its start and stop dis-
tance.
See also CALCulate<Chn>:TRANsform:DTFault:CENTer and
CALCulate<Chn>:TRANsform:DTFault:SPAN.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Start>/<Stop> Start/stop distance.
Default unit: m
Example:
:CALCulate1:TRANsform:DTFault:STARt 1; STOP 3
Defines the DtF distance window of the active trace of channel 1
as [1m, 3m].
Options: R&S ZNB/ZNBT-K3
Manual operation: See "Start Distance / Stop Distance" on page 678
CALCulate<Chn>:TRANsform:DTFault:STATe <DtfState>
If channel <Chn> is performing an unsegmented frequency sweep, this command
allows to transform its active trace into a Distance to Fault (DtF) trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<DtfState> OFF|ON – DtF disabled|enabled
*RST: OFF
SCPI Command Reference

Table of Contents

Other manuals for Rohde & Schwarz ZNB

Related product manuals