Command Reference
R&S
®
ZNB/ZNBT
965User Manual 1173.9163.02 ─ 54
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<Inductance> Inductance L for ground loop deembedding.
Range: -1 H to 1 H.
Increment: 1 pH (1E-12 H)
*RST: 1 nH (1E-9 H)
Default unit: H
Parameters for setting and query:
<CircuitModel> SL
Possible circuit models (character data); see Table 8-8.
Example:
*RST; :CALC:TRAN:VNET:GLO:DEEM:PAR:L? SL
Query the default inductance for ground loop deembedding. The
response is 1E-009 (1 nH).
CALC:TRAN:VNET:GLO:DEEM:PAR:L SL, 2.2E-9
Increase the inductance to 2.2 nH.
Manual operation: See "Network" on page 610
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:R <CircuitModel>, <Resistance>
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:
PARameters:R? <CircuitModel>
Specifies the resistance value R in the different circuit models for ground loop deem-
bedding.
Suffix:
<Ch>
.
Channel number.
<group> Port group (DUT) number.
If multiple port groups are configured (see SOURce<Ch>:
GROup<Grp>:PPORts) and CALCulate<Ch>:TRANsform:
VNETworks:GLOop:GROup is ON, then each port group can
have its own de-/embedding models.
Parameters:
<Resistance> Resistance R for the specified circuit model.
Range: -10 MΩ to 10 MΩ
Increment: 1 mΩ
*RST: 0 Ω if the resistance is connected in series with an
inductance. 10 MΩ if the resistance is connected in
parallel with a capacitance)
Default unit: Ω
SCPI Command Reference