1018User Manual 1173.9163.02 ─ 54
Manual operation: See "Range / Shunt" on page 520
2 μA 20 μA 200 μA 2 mA 20 mA
Shunt 10 Ω 100 Ω 1 kΩ 10 kΩ 100 kΩ
CONTrol:RFFE<Bus>:TEST:DATA:VOLTage?
CONTrol:RFFE<Bus>:TEST:CLOCk:VOLTage?
CONTrol:RFFE<Bus>:TEST:VIO:VOLTage?
Returns the results of the voltage measurement on the related RFFE pin.
Suffix:
<Bus>
.
RFFE bus
Return values:
<Measured> Measured voltage
Usage: Query only
Options: R&S ZN-B15/Z15 Var. 03
Manual operation: See "Voltage, Current" on page 521
CONTrol:RFFE<Bus>:TEST:DATA:CURRent?
CONTrol:RFFE<Bus>:TEST:CLOCk:CURRent?
CONTrol:RFFE<Bus>:TEST:VIO:CURRent?
Returns the results of the current measurement on the related RFFE pin.
Suffix:
<Bus>
.
RFFE bus
Return values:
<Measured> Measured current
Usage: Query only
Options: R&S ZN-B15/Z15 Var. 03
Manual operation: See "Voltage, Current" on page 521
CONTrol<Ch>:SEGMent<Nr>:SEQuence:COUNt?
For segmented sweeps this command queries the length of the command/switch
sequence defined for a particular channel and sweep segment.
The command/switch sequences are defined using CONTrol<Ch>:SEGMent<Nr>:
SEQuence<Nr>:RFFE<Bus>:COMMand:DATA and CONTrol<Ch>:SEGMent<Nr>:
SEQuence<Nr>:GPIO<Port>:VOLTage.
Suffix:
<Ch>
.
Channel number
<Nr> Segment number